Maintenance and precautions of centrifuges
作者:時(shí)間:2017/6/24 15:58:27閱讀:1202
1. The installation of the centrifuge should be based on the performance of the centrifuge to select a suitable location for placement. For general low-speed centrifuges, they can be placed on a stable and solid tabletop. The bases are all equipped with rubber suction feet. With the help of high pneumatic pressure and the weight of the instrument itself, they are closely attached to the tabletop.
2. Large-capacity low-speed centrifuges, high-speed centrifuges and high-speed refrigerated centrifuges should be placed on a solid ground and placed horizontally accordingly. Before choosing a suitable rotor centrifuge to work, the load should be balanced. The smaller the weight error, the better. Otherwise, it will cause severe vibration and damage the centrifugal rotor and shaft.
3. Centrifuge stand: The centrifuge separation stand must be operated after the centrifuge tubes are placed. It is strictly prohibited to run it empty.
2. Large-capacity low-speed centrifuges, high-speed centrifuges and high-speed refrigerated centrifuges should be placed on a solid ground and placed horizontally accordingly. Before choosing a suitable rotor centrifuge to work, the load should be balanced. The smaller the weight error, the better. Otherwise, it will cause severe vibration and damage the centrifugal rotor and shaft.
3. Centrifuge stand: The centrifuge separation stand must be operated after the centrifuge tubes are placed. It is strictly prohibited to run it empty.